Metrant
Humming V is done (Figure ; That humming V is done outward appearance photograph)
1. The characteristic
As for that MERA humming II is done, with the manufacture of the "METORAN" Inc., the Izumi carpenter medical department industry became a sales agent. Later, that humming II is done became a discontinuance, that humming V is done was developed by the "METORAN" Inc. and as for this, Daido Hoxan became a sales agent. At present, Iwaki Inc. is selling. Powerful neoveto rival product however, there is that humming V is done. That humming V is done, too, was a resuscitator for the domestic newborn baby who combined HFO of the piston method and IMV of the continuous flow relieving method, but the stroke volume increased more to 50 ml than humming II is done and the application range expanded. Also, because MPU had two ones of 16 bits, the resolution of the control improved. That humming V is done has a pressure trigger mechanism but the trigger performance in the newborn baby isn't enough. The HFO adaptation is a maximum of about 8 kg at the weight. R100 which is Calliope alpha where the "METORAN" Inc. updated a that humming V is done mechanism in the modern technology after that and where there was cost cutting and HFO machine for the infant adult, too, is •Ή”„ing. The thing with schedule that improvement is repeated every model about the HFO mechanism of the "METORAN" Inc. and that a model is unified by the X of next term main force aircraft's being hummed in the future
2. The performance
1) The mode that it is possible to use
(All active-ventilations are done by the pressure relieving ventilation of time cycle.)
(S) IMV, the pressure relieving
CPAP (The continuous flow)
HFO
HFO+AutoSigh
2) This proximal dater
The continuous flow quantity .5-45 LPM (The IMV mode)
.8 LPM (The HFO mode)
The intake gas slew-rate L/s2
The IMV number of times .1-120 BPM
The HFO number of times .13-17 Hz
The maximum HFO amplitude of vibration .80 ml
3. The explanation of the control circuit, the controlling mechanism
The method which the ventilating-pressure is composed of by relieving a continuous flow in the pressure by the exhalation-valve in the time cycle is adopted. Call feeling ‘ͺ is equipped with HFO of the piston type. Because the control of flow does, that it is mechanical, the new Mattick circuit is complicated.
4. The new Mattick circuit (Figure ; That humming V is done new Mattick circuit)
1) The IMV/CPAP mode (Figure ; Outline of gas stream in IMV/CPAP mode)
The dense fog which was entered than each gas of oxygen, air supplies an intake flow through positive-crankcase-ventilation-valve (1)EP1 and (2)EP2 after adjusted to the setting oxygen concentration by the isotonicity mechanical blender. At this time, current-adjustment (1) valve EP1 does a current-adjustment in case of intake and (2)EP2 adjusts a continuous flow through the reservoir back. Gas is sent to the patient circuit, but the safety valve operates and being opened to it in the atmosphere when the patient circuit internal pressure exceeds 15cmH2O than the caution set value Gas is sent to the lung through the impedance valve. At this time, the impedance valve becomes the condition (the influence doesn't come out to the intake flow) which the disengagedness and the becoming resistance don't hang over. An expiratory-pressure is adjusted to the gas which was summoned from the lung by the exhalation-valve through the circuit of breathing of the side of the expiration and is opened to it in the atmosphere.
2) The HFO mode (Figure ; Outline of gas stream in HFO mode)
The dense fog which was entered than each gas of oxygen, air supplies intake flow 8L/a minute through positive-crankcase-ventilation-valve (1)EP1 after adjusted to the setting oxygen concentration by the isotonicity mechanical blender. At this time, positive-crankcase-ventilation-valve (1)EP1 adjusts a continuous flow in case of intake. Positive-crankcase-ventilation-valve (2)EP2 adjusts a continuous flow to control an average airway pressure through venture Lee. The piston adds a high frequency oscillation for the HFO ventilation into breathing circuit with the circuit of the side of the piston. Gas is sent to the patient circuit, but the safety valve operates and being opened to it in the atmosphere when the patient circuit internal pressure exceeds 15cmH2O than the caution set value Gas is sent to the lung through the impedance valve. At this time, the impedance valve becomes a half •Β condition and the condition (the continuous flow make pass and cuts a high frequency oscillation ingredient) which the becoming resistance drives. An average airway pressure is adjusted to the gas which was summoned from the lung by the exhalation-valve through the circuit of breathing of the side of the expiration with the continuous flow in 8L/the minute and is opened to it in the atmosphere. It is removed, it is sent to the piston integrate type exhalation-valve and moisture and so on are opened to the expiration which was emitted from the patient about any ventilation mode in the atmosphere after that by the watered lap. In the exhalation-valve, the timing of intake and the expiration resembles each other with the control unit, opens and shuts and it controls an average airway pressure at the time of the PEEP/CPAP pressure and the HFO ventilation mode.
3) The structure of the HFO piston (Figure ; Photograph of V piston)
The that humming V is done vibration generating procedure adopts a piston method. The structure can be roughly divided into the linear motor to move this piston, and the control unit, the new Mattick control unit which manages ventilator ability. The gas which plumbing was inputted to passes a decompression-valve and an oxygen blender, it is controlled and the side of intake is sent to breathing circuit in the flow rate by the flow valve. The flow valve implements the one which is special respectively for the IMV/CPAP mode and the HFO mode. The side of the expiration is separated to two, and as for the side, the exhalation-valve and jet venture Lee are implemented and get off. The low-pass filter tube, the other which is thin to control the emission of the expiration and the continuous flow and average airway pressure (MAP) become the piston tube which is bold to tell the oscillating flow to have made occur with the piston to breathing circuit from the side of the expiration. In the IMV/CPAP mode, a flow inspiratory-pressure, PEEP/CPAP pressure and so on are controlled in the expiration by the low-pass filter tube. The piston tube gets not to move with a piston shut for the leak to pass away. In the HFO mode, the control of MAP is made a low-pass filter tube by the emission of the flow continuous flow, the expiration and jet venture Lee in the expiration. The piston vibrates at the stroke volume (SV) to have been opened to and for it to have been set by the linear motor and frequency (Hz) and vibration is conducted into breathing circuit by the piston tube.
5. The control software
@
1) The trigger formula
It is a pressure trigger formula.
2)IMV/CPAP
SIMV is a variable-time method. (Figure ; Explanation of SIMV : The length of t5, t6, t7, t8 is the same and is fixed at the 60sec/SIMV number of times.)
3) In SIGH addition, HFO stops. (Figure ; Explanation of auto sigh) The setting of IMV is used in the number of times and the time of SIGH. That is, the operation of SIGH is fixed at the air changes per hour and the intake time by IMV.
6. The operation system (Figure ; That humming V is done operating panel)
In the IMV mode, it sets trigger sensitivity, an inspiratory-pressure, a flow rate, intake time, breathing number of times. In the HFO mode, it sets SIGH pressure, a frequency, a stroke volume, an average airway pressure. When using SIGH, it sets a desired air changes per hour (the SIGH number of times) and intake time (the SIGH time) by the setting of IMV.
7. The monitor, the alarm function
An airway pressure is displayed in the circular liquid crystal bargraph. Because Fluothane Sir isn't had, the taking air quantity of ventilation and the expired volume per minute and so on can not be monitored. There is an alarm of the high airway pressure, the low airway pressure, the PEEP/CPAP pressure malfunction, being power failure, the motor abnormality, the oxygen-tension decline, the air-failure.
8. The display feature
It puts on an average airway pressure or amplitude, an average airway pressure or SIGH pressure, PEEP/CPAP pressure or a frequency by the numeral display.
9. The patient circuit composition (Figure ; That humming V is done patient circuit)
It was shown in the figure.
10. The daily maintenance
It washes a piston and an impedance valve sufficiently at the enzyme system detergent. After that, it rinses sufficiently so as not for a detergent to be left. After making dry up fast, it sterilizes in EOG. The piston can be an autoclave, too.
11. The regular inspection
It does piston maintenance every 6 months. It does regularly inspecting, an overhaul in 2 every years.
12. The fault
1) The patient circuit of call feeling ‘ͺ is complicated.
2) The management of the piston is complicated.
3) The monitor of the air quantity of ventilation isn't made.
4) PTV isn't made.
@
Calliope alpha (Figure ; Outward appearance photograph of Calliope alpha)
1. The characteristic
It is Calliope that improved a that humming V is done mechanism modernistically. In that humming V is done, it was CPAP method by the continuous flow + reservoir bag but Calliope advanced on the ventilating-pressure control system of the demand type by PSV with the base flow who can express with the continuous flow, too. On the side of intake, Fluothane Sir is had but it is not in call feeling ‘ͺ. The gas flow rate controllability, too, improved and a piston, too, was changed into the level type. The adaptation of HFO is a maximum of about 10 kg at the weight.
2. The performance
1) The mode that it is possible to use
(S) IMV(PC)+PSV
CPAP (The continuous flow + demand)
HFO
2) This proximal dater
The continuous flow quantity .1-20 LPM (The time of trigger off)
3-6LPM (The time of trigger on)
The highest value for expiratory flow .120LPM
The IMV number of times .1-120 BPM
The HFO number of times .13-17 Hz
The maximum HFO amplitude of vibration .80 ml
The HFO frequency .15Hz
3. The explanation of the control circuit, the controlling mechanism
It is done using the servo valve by the combination of the ironhand "HA" intake valve and the expiration valve. Therefore, the mechanism could be simplified when comparing with that humming V is done and could have measured cost cutting, too. The HFO frequency, too, is a fixation. For the pressure trigger formula, PTV of the newborn baby isn't made. SIMV however, the trigger can not be gotten when floating down a lot of continuous flows.
4. The new Mattick circuit (Figure ; block diagram of Calliope alpha) (Figure ; New Mattick schematic of Calliope alpha)
The entered dense fog enters from each gas of oxygen, air to the filter and the decompression-valve respectively. The ratio of oxygen and air is fixed to become the inhalation oxygen concentration which was set with the panel and this ratio is kept regardless of the flow rate.
1) The IMV/CPAP ventilation mode
Airway pressure control during intake is done at both of the exhalation-valve aperture and the intake flow. In the expiration, it is a panel, and the constant flow with set flow rate flows and airway pressure control is done by the exhalation-valve aperture. The safety valve operates and opened to it in the atmosphere when the patient circuit internal pressure exceeds 15cmH2O than the caution set value Gas is sent to the lung through the H/L valve (the impedance valve). At this time, the impedance valve becomes the condition (the influence doesn't come out to the intake flow) which the disengagedness and the becoming resistance don't hang over. An expiratory-pressure is adjusted to the gas which was summoned from the lung by the exhalation-valve through the expiration line and is opened to it in the atmosphere.
2) The HFO ventilation mode
MAP control is done at both of the exhalation-valve aperture and the intake flow but the intake flow never falls below 8L/the minute. Intake gas is sent to the lung through the impedance valve. At this time, the impedance valve becomes a half •Β condition and the condition (the continuous flow make pass and cuts a high frequency oscillation ingredient) which the becoming resistance drives. The piston adds a high frequency oscillation for the HFO ventilation into breathing circuit with the HFO vibration line. The safety valve operates and opened to it in the atmosphere when the patient circuit internal pressure exceeds 15cmH2O than the caution set value while ventilating An average airway pressure is adjusted to the gas which was summoned from the lung by the exhalation-valve through the circuit of breathing of the side of the expiration with the intake gas stream and is opened to it in the atmosphere.
5. The control software
1) The trigger formula
It is a pressure trigger formula. The maximum sensitivity is - 0.5cmH2O. The sensitivity declines relatively when floating down a lot of continuous flows (setting with the control with "flow rate") and the trigger can not be caught.
2)PSV
The expiration recognition condition of PSV can be set in 10-50% of ranges. When turning a control with "flow rate" while making SIMV mode and pressing "the display change" of "the pressure monitor", the display switches over to the expiration recognition and becomes able to be set.
3)SIMV
It is the variable-time method which is the same as that humming V is done.
6. The operation system (Figure ; Operating panel photograph of Calliope alpha)
It sets necessary items such as the air changes per hour and the ventilating-pressure with the control. The setting of an expiration recognition condition and Rhys time switches over to each setting when it turns a control in "flow rate" "the intake time" while it presses "the display change". The others are the same as that humming V is done approximately.
7. The monitor, the alarm function
It is the same as that humming V is done approximately.
8. The display feature
It is the same as that humming V is done approximately.
9. The patient circuit composition (Figure ; Patient circuit of Calliope alpha)
It was shown in the figure. It is the same as that humming V is done approximately.
10. The daily maintenance
It is the same as that humming V is done approximately.
11. The regular inspection
It is the same as that humming V is done approximately.
12. The fault
1)PTV
It doesn't put on with the trigger in case of not being a flow trigger formula and continuous flow's there being for the pressure trigger formula. Therefore, PTV of the newborn baby is impossible. It is difficult to understand to adopt a pressure trigger though the continuous flow is a necessary mechanism. Therefore, basically, the trigger mechanism has uprooted "a free gift". PSV, too, can not work neatly if the trigger performance is enough.
2) Fluothane Sir
That it isn't possible to do the measurement of the air quantity of ventilation while there is Fluothane Sir specially is sorry. It is in one of the cause that there is not Fluothane Sir in call feeling ‘ͺ.
3) Rhys time
The Rhys time adjustment and the expiration recognition condition can be changed but it is question about whether or not to set no matter how appropriately though there is not a graphic monitor, too.
4) The concept of the design will are not only that humming V is done cost cutting but that the precious improvement doesn't live sufficiently as the improvement is sorry.